1887

n South African Journal of Child Health - Point-of-care estimation of haemoglobin concentration in neonates and infants

USD

 

Abstract

The HemoCue is a point-of-care analytical system for haemoglobin concentration (Hb) measurement. Point-of-care testing has been validated in hospitals and outpatient departments to assist with urgent patient management by providing rapid laboratory test results.


In this prospective study we compared the analytical performance of the HemoCue with that of the Advia 120 haematology analyser with regard to accuracy, precision and linearity in the measurement of Hb in neonates and infants.
Samples from 44 patients were analysed by both instruments and the results compared using difference plots. The mean Hb value for the HemoCue (11.8 g/dl; range 4.8 - 18.7) was comparable to that for the Advia (11.8 g/dl; range 5.2 - 19.2). The Bland-Altman difference plot revealed good agreement. Bias between the two methods was small and the imprecision was within acceptable limits. Hb measurement was linear in the range 4.8 - 20 g/dl.
In neonates and infants, the diagnostic accuracy of the HemoCue point-of-care device is comparable with that of the Advia 120 analyser. In neonatal and paediatric units, where the volume of blood available is limited and turnaround time critical, an appropriately quality-assured HemoCue can replace standard haematology analysers in the measurement of Hb.

Loading

Article metrics loading...

/content/m_sajch/6/1/EJC64842
2012-03-01
2016-12-05
This is a required field
Please enter a valid email address
Approval was a Success
Invalid data
An Error Occurred
Approval was partially successful, following selected items could not be processed due to error