n African Vision and Eye Health - Repeatability and reproducibility of horizontal corneal diameter and anterior corneal power measurements using the Oculus Keratograph 4 : original research

Volume 75 Number 1
  • ISSN : 2413-3183
  • E-ISSN: 2410-1516
This is currently unavailable for purchase.

 

Abstract

To evaluate the repeatability and reproducibility of horizontal corneal diameter (HCD) and anterior corneal power (ACP) measurements obtained with the Oculus Keratograph 4 (OCULUS Optikgeräte GmbH).


These parameters (HCD and ACP) were prospectively measured in quick succession three times in each of the right eyes of 40 healthy subjects, aged 18-28 years, with normal vision (6/6 or better visual acuity) in the first session by a single examiner. Measurements were then repeated in the second session scheduled 1 week later by the same examiner using the same instrument. Repeatability and reproducibility of HCD and ACP measurements was assessed based on the intra-session and intersession within-subject standard deviation (s), repeatability (2.77s), coefficient of variation (CoV) and intra-class correlation coefficient (ICC).
Intra-session repeatability and intersession reproducibility of all measured parameters showed a repeatability (2.77s) of 0.35 mm or less for HCD and 0.35 D or less for ACP, a CoV of 0.30% or less and an ICC of more than 0.9.
HCD and ACP measurements obtained using an Oculus Keratograph 4 show good repeatability and reproducibility in healthy eyes; therefore, these parameters can be used for longitudinal follow-up when measured with this device.

Loading full text...

Full text loading...

Loading

Article metrics loading...

/content/aveh/75/1/EJC195003
2016-01-01
2019-08-25

This is a required field
Please enter a valid email address
Approval was a Success
Invalid data
An Error Occurred
Approval was partially successful, following selected items could not be processed due to error